Philips EverFlo Q Service And Technical Reference Manual page 60

Oxygen concentrator
Hide thumbs Also See for EverFlo Q:
Table of Contents

Advertisement

To remove the Sieve Assembly
1. Remove the Filter Cover. Refer to the Filter Cover Replacement section for more information.
2. Remove the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
3. Remove the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
4. Remove the Pressure Regulator Adaptor. Refer to the O2 Quick Coupler Replacement section for
more information.
5. Remove the Valve Solenoid connector from the J7 location of the Main PCA.
6. Remove the Purge Valve Connector from the J8 location of the Main PCA (230V devices equipped
with a Purge Valve Only).
Figure 8-17: PURGE VALVE LOCATION ON SIEVE CANISTER ASSEMBLY
7. Disconnect the blue hose going from the Compressor to the Sieve Canister Assembly.
8. Lift the Sieve Canister Assembly up and out of the slots located in the Front Cabinet.
Page 60
Figure 8-18: SIEVE CANISTER ASSEMBLY REMOVAL
LO TM
E
F
S
VER
ERVICE
& T
I
ECHNICAL
NFORMATION
1039055, VER. 06