To install the Pressure Regulator Assembly
1. Using pliers, clamp the flow tubing to the Pressure Regulator Assembly.
2. Place the Pressure Regulator Assembly in place by lining the Pressure Regulator Adaptor up with
the grooves in the Sieve Canister Assembly.
3. Lock the Pressure Regulator Assembly in place by turning the Assembly downwards.
Figure 8-13: INSTALLING THE PRESSURE REGULATOR
4. Install the clear tubing to the O
Coupler and secure with a tie wrap.
2
5. Route the clear tubing under the Sieve Canister Assembly.