9. Remove the Solenoid Valve Assembly from the Sieve Canister Assembly by removing the five phil-
lips head screws.
To install the Sieve Assembly
1. Install the Solenoid Valve Assembly. Refer to the Valve Solenoid Replacement section for more
information.
2. Side the Sieve Canister Assembly into the hole in the bottom of the Front Cabinet.
3. Place the top and bottom of the Sieve Canister Assembly into the guide slots in the Front Cabinet.
4. Connect the blue hose from the Compressor to the Sieve Canister Assembly.
5. Connect the Valve Solenoid connector to J7 of the Main PCA.
6. Connect the Purge Valve connector to J8 of the Main PCA (230V devices equipped with a Purge
Valve Only).
7. Install the Pressure Regulator Assembly. Refer to the O2 Quick Coupler Replacement section for
more information.
8. Install the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
9. Install the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
10. Install the Filter Cover. Refer to the Filter Cover Replacement section for more information.
11. Install the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
Devices manufactured after 06/18/2013 will be equipped with longer sieve canisters (P/Ns
1050248 & 1052382). Unless the unit is a Trans Fill capable or International model, the unit will
use P/N 1050248; otherwise, P/N 1052382 is used. Older devices will be equipped with shorter
sieve canisters (P/N 1038825).
Page 61
Figure 8-19: SLOT LOCATIONS
LO TM
E
F
S
& T
VER
ERVICE
NOTE
I
ECHNICAL
NFORMATION
1039055, VER. 06