SYMPTOM
Fluctuations in
oxygen pressure
Fluctuation in flow
ball
Unit will not turn on
Page 34
CAUSE
Pressure Regulator
failure
Valve Solenoid failure
No power to the unit
LO TM
E
F
S
VER
VERIFICATION
Verify the flow from
regulator is correct
Perform Outlet /
Regulated Pressure
Test
Verify the Valve
Solenoid is working
properly
Verify the valve
wiring harness is
connected at the
Main PCA
Verify the power cord
is connected to the
wall outlet
Verify there is power
in the wall outlet
Verify if the outlet is
connected to a light
switch and the switch
is in the ON position
& T
I
ERVICE
ECHNICAL
NFORMATION
CORRECTIVE
ACTION
Replace the
regulator
Replace the Valve
Body
Reconnect the
wiring harnesses
and retest
If problem
continues, install
old valve body
and new pilot
l
Connect the
power cord to the
wall outlet
Check the
household fuse or
circuit breaker
Move the power
switch to the ON
position
1039055, VER. 06