Schematic of CI flow
control module
Flow module state diagram:
Result
Gas A flows
Control panel lights (LEDs)
Gas A (green)
On
Gas B (amber) Off
Purge (red)
Off
Gas Off (red)
Off
Valve state
Valve A
Open
Valve B
Closed
On → setpoint
MFC
Isolation valve Open
Gas A
(Methane)
supply
Gas select
valve A
Gas select
valve B
Gas B
(Other gas)
supply
Gas B flows
Purge
with Gas A
Off
On
On
Off
Off
On
Off
Off
Closed
Open
Open
Closed
On → setpoint
On → 100%
Open
Open
Green
light
Mass
Flow
Controller
Calibration
valve
Amber
light
Restrictor
PFDTD
vial
Purge
with Gas B
Off
On
On
Off
Closed
Open
On → 100%
Open
 Troubleshooting
$LU OHDNV
CI ion
source
Isolation
valve
EI/CI
GC/MSD
interface
GC column
Pump out
Standby, vented,
flow module
or EI mode
Off
Off
Off
Off
On
Off
On
On
Closed
Closed
Closed
Closed
On → 100%
Off (→0%)
Open
Closed