To replace the methane/isobutane gas purifier

Materials needed:
Methane/isobutane gas purifier (G1999-80410)
Front ferrule for 1/8-inch tubing (5180-4110)
Rear ferrule for 1/8-inch tubing (5180-4116)
Tubing cutter (8710-1709)
The methane/isobutane gas purifier needs to be replaced after four tanks of reagent
gas. This frequency may vary depending on purity of the gas and care taken in
uncapping and installing the gas purifier. A large leak upstream from the gas purifier
can quickly exhaust the reduced metal of its oxygen and moisture traps.
1 To install the methane/isobutane gas purifier, follow the instructions on
the label for installation and replacement intervals.
C A U T I O N
Be sure not to remove the caps until you are ready to install the gas purifier. Only remove
the caps in the gas flow to prevent contamination by air.
0HWKDQH LV IODPPDEOH ([WLQJXLVK DOO IODPHV LQ WKH DUHD EHIRUH WXUQLQJ RQ JDV IORZ
W A R N I N G
2 Disconnect the fittings on the old filter.
3 Remove the ferrules from the tubing at the outlet of the gas purifier.
Using the tubing cutter, cut off the end of the tubing with the ferrules.
4 Install the new filter.
5 Purge the new filter.
6 Cap the old filter and prepare it to be sent for regeneration.
See the instructions on the label.
7R UHSODFH WKH PHWKDQHLVREXWDQH JDV SXULILHU
 Maintenance