Maintenance - Philips ReCare 7.0 R User Manual

Physical therapy solutions
Table of Contents

Advertisement

Maintenance

• Wipe down all areas in the sweat path with a damp cloth after each
use to prevent rust.
• Check the pedal to make sure they are tight (monthly).
• If a squeak, thump, clicking or rough feeling develops the main
cause is most likely one of two reasons:
- The hardware was not su ciently tightened during assembly. All
bolts that were installed during assembly need to be tightened as
much as possible. It may be necessary to use a larger wrench than
the one provided if you cannot tighten the bolts su ciently. I
cannot stress this point enough; 90% of calls to the service
department for noise issues can be traced to loose hardware.
- The crank arm nut and/or the pedals need to be retightened.
• If squeaks or other noises persist, check that the unit is properly
leveled. There are 2 leveling pads on the bottom of the rear
stabilizer, use a ½" wrench (or adjustable wrench) to adjust the
levelers.
Maintenance menu in console software
The console has built in maintenance/diagnostic software. The software
will allow you to change the console settings from English to Metric and
turn o the beeping of the speaker when a key is pressed for example. To
enter the Maintenance menu (may be called Engineering mode, depend-
ing on version) press and hold down the Start, Stop and Enter keys. Keep
holding the keys down for about 5 seconds and the message window will
display "Maintenance mode". Press the enter button to access the menu
below:
64

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents