Guidance and manufacturer's declaration –
electromagnetic compatibility
The 7.0 S is intended for use in the electromagnetic environment speci ed below. The customer or
the user of the 7.0 S should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Immunity test
IEC 60601
test level
Electrostatic
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±8 kV air
Power
frequency
(50/60 Hz)
3 A/m
magnetic eld
IEC 61000-4-8
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
Compliance
Group 1
The 7.0 S uses
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
Class B
The 7.0 S is suitable for use in all
establishments, including domestic
establishments
Compliance level
±6 kV contact
Floors should be wood, concrete or
ceramic tile. If oors are covered with
±8 kV air
synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic elds
should be at levels characteristic of a
3 A/m
typical location in a typical
commercial or hospital environment.
d = 1,2
3 V/m
d = 2,3
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
63
Electromagnetic environment –
guidance
RF energy only for
Electromagnetic environment
guidance
80 MHz to 800 MHz