HP 16533A Programmer's Manual page 31

1-gsa/s and 2-gsa/s digitizing oscilloscope
Hide thumbs Also See for 16533A:
Table of Contents

Advertisement

Figure 3-1
ACQuire Subsystem Syntax Diagram
Table 3-1
ACQuire Parameter Values
Parameter
count_arg
Acquisition Type Normal
In the Normal mode, with the ACCumulate command OFF, the oscilloscope
acquires waveform data and then displays the waveform. When the
oscilloscope makes a new acquisition, the previously acquired waveform is
erased from the display and replaced by the newly acquired waveform. When
the ACCumulate command is ON, the oscilloscope displays all the waveform
acquisitions without erasing the previously acquired waveform.
Acquisition Type Average
In the Average mode, the oscilloscope averages the data points on the
waveform with previously acquired data. Averaging helps eliminate random
noise from the displayed waveform. In this mode the ACCumulate command
is OFF. When Average mode is selected, the number of averages must also
be specified using the COUNt command. Previously averaged waveform data
is erased from the display and the newly averaged waveform is displayed.
ACQuire Subsystem
Value
An integer that specifies the
number of averages to be taken of
each time point. The choices are
2, 4, 8, 16, 32, 64, 128, or 256.
3–3

Advertisement

Table of Contents
loading

This manual is also suitable for:

16534a