Factory and acceleration settings
Enter the Factory settings; press Start and Speed Fast keys while
console is in a.power up reset. User presses enter
• Units: English
- The default setting is English. User can press any up/down arrows
to change to Metric. User presses enter.
• Adjust min speed then press enter
- Default value is 0.1 mph and can be adjusted up to 0.5 mph. The
speed number to be shown in the speed window.
• Adjust max speed then press enter
- Default value is 10.0 mph and can be adjusted down to 3.0 mph.
The speed number to be shown in the speed window
• Adjust reverse speed max then press enter
- Default value is 3.0 mph and can be adjusted down to 1.0 mph and up
to 5.0 mph.
• Adjust belt speed acceleration time per each 1.0 mph
- The default is 0:03 seconds and will be shown in the Time window.
The time can be adjusted down to 0:01 and up to 1:00 minute
• Adjust belt speed deceleration time per each 1.0 mph
- The default is 0:03 seconds and will be shown in the Time window.
The time can be adjusted down to 0:01 and up to 1:00 minute
• Press enter for grade calibration or to exit hold stop key until
reset.
• Adjust max incline then press enter
- The default value is 15 and can be adjusted down to 10
• Adjust max decline then press start to calibrate
- The default value is 5 and can be adjusted up to 10
• If Grade calibration is successful the MW shows "passed" for 3
seconds then exit to idle mode.