Immunity test
Electrical fast
+/-2 kV for power
transient/burst
supply lines
IEC 61000-4-4
+/-1 kV for
input/output lines
!
!
Note
• If the device is interfered by power or signal cable, image quality
may be reduced or abnormally displayed. Such kind of interference
images could be easily identi ed and di erentiated from the physi-
ological characteristics of patient and longer clinical time consumed
but wouldn't have any diagnostic accuracy issue.
• If there is a certain frequency of image interference, there is a need
of isolation or ltering of the RF signal.
IEC 60601
Compliance level
test level
+/-2 kV for power
supply lines
+/-1 kV for
input/output lines
!
70
Electromagnetic environment
guidance
Mains power quality should be that
of a typical commercial or hospital
environment.
!
–