Log Type; Starttime Prompt - GE DigitalFlow GF868 Programming Manual

Ultrasonic flowmeter for flare gas
Hide thumbs Also See for DigitalFlow GF868:
Table of Contents

Advertisement

3.5
Creating an ERROR Log (cont.)
In addition, refer to Chapter 3, Diagnostics, of the Service Manual for a discussion of the many parameters accessible
DIAG
via the
option.
Note: The units assigned to the parameters in Table 17 on page 58 are those selected in the User Program (
submenu).
6. Use the [, [ and
7. Use the [, [ and
If you selected a totalized value for display, go to step 8.
If you did not, go to step 9.
8. At the Set Log Totals to 0 prompt, press
totals to zero. Choosing

3.5.1 Log Type

[F1]
9. Press
to create a non-circular log or press
A non-circular log automatically stops when the meter runs out of memory or the specified
circular log runs continuously until manually stopped, but only the data from the most recent log cycle is saved. At the
start of each cycle, the oldest recorded data is overwritten by the new data from the current cycle.
CAUTION! If data from one circular log cycle exceeds the meter's memory capacity, the
earliest logged data will be lost.

3.5.2 STARTTIME Prompt

[F1]
10. Press
to accept the displayed start time or press
[F3]
press
.
OK
If
was selected, proceed to step 11.
NOW
If
was selected, you have finished setting up the error log. Proceed to Procedure Options on page 60.
EDIT
If
was selected, complete the steps listed on the next page.
[F1]-[F2]
a. Press
to select
earlier than the current time will generate an error message.)
b. Enter the desired Minutes and press
c. Enter the desired Seconds and press
DigitalFlow™ GF868 Programming Manual (2-Channel)
[F1]-[F4]
keys to select the second parameter to be logged.
[F1]-[F4]
keys to select the third parameter to be logged.
[F1]
to leave the log totals at their present value or press
YES
at this prompt only resets the log totals to zero and has no effect on the meter totals.
[F2]
AM
PM
or
. Then, enter the desired Hour (1-12) and press
[ENT]
. The acceptable range is 0 to 59.
[ENT]
. The acceptable range is 0 to 59.
to create a circular log.
[F2]
to enter a different start time. To start logging immediately,
Chapter 3. Logging Data
SYSTM
[F2]
to reset the log
END TIME
is reached. A
[ENT]
. (Entry of a start time
59