...10 PROGRAMMING

10.3 Range 3

10.3.1 Access to Secure Parameters (LOWER DISPLAY) – AK101 only
A 5-digit code is used to prevent unauthorized access to the secure parameters.
SECURITY CODE
10.3.2 Language Page
English
10.3.3 Set Up Outputs
SET UP OUTPUTS
Alarm H2 IN *
AIR IN *
*Note. Purge gas options include:
CO
(Carbon dioxide)
2
N
(Nitrogen)
2
Ar
(Argon)
40
Security Code
Enter the required code number, between 00000 and 19999, to gain access to the
secure parameters. If an incorrect value is entered, access to subsequent programming
pages is prevented and the display reverts to the
Note. The security code is preset at '00000' to allow access during commissioning but
should be altered to a unique value, known only to authorized operators – see
parameter in
Security Code
Advance to
Language Page
Language Page
Select the required language for the display.
Advance to
Set Up Outputs Page
Advance to next parameter.
Purge Gas Alarm Select
Select the parameter on which the purge gas alarm is to operate.
Reprogram as required to agree with selected range:
for Range 2 select H2 IN PURGE GAS*,
for Range 3 select AIR IN PURGE GAS*.
Continued on next page.
Operating Page
.
Set Up Outputs Page
.
.
.
Alter