Performance Tests
STM-0/STS-1 Transmitter Output Waveshape
STM-0/STS-1 Transmitter Output Waveshape
Specifications
Level
Cross-Connect
530 mV peak (nominal). Compliant with GR-253, Issue
(450 feet)
1, Figure 4-10 and draft recommendation ITU-R F.750,
Appendix1
HIGH
1 V peak (nominal)
LOW
300 mV peak (nominal)
Description
An Oscilloscope is connected to the HP 37718A/19A Transmitter STM-0/
STS-1 output and used to check the STM-0/STS-1 waveshape against the
relevant mask.
Equipment Required
Oscilloscope
Ω
75
Termination
T Connector
Procedure
Positive Pulse
1 Recall HP 37718A/19A DEFAULT SETTINGS as shown on page 3-4
and set up the
TRANSMIT
3-66
Pulse Amplitude & Shape
: HP 54520A Option 001
: HP 15522-80010
: HP 1250-0781
display as shown on the following page.