Performance Tests
Table 3-1 Recommended Test Equipment
Instrument
Oscilloscope
Frequency Synthesizer 75Ω Output, Sinewave to 80 MHz,
Signal Generator
Frequency Counter
Spectrum Analyzer**
Optical Coupler**
Optical Power Meter
and Sensor Module
Lightwave Converter
Optical Attenuator
FC/PC Optical
Interface Connector
Optical Cables
PDH Structured Test
Set
64 kb/s Clock
Generator
75Ω Attenuator Kit
Cable Simulator #E1
Cable Simulator #E2
Cable Simulator #E3
Cable Simulator #E4
3-2
Critical Specification
400 MHz Bandwidth, 1 MΩ Input Termination
Telecom mask measurement capability.
Amplitude to 3 V pk-pk at 10 kHz, 1 Hz resolution
Sinewave 700 kHz to 170 MHz, Amplitude 500 mV
Range 0 to 200 MHz, 2 channels with accuracy
<0.1ppm. (Ratio Mode)
Frequency >2.5 Gb/s. Resolution bandwidth 10 Hz.
1310 and 1510 nm. 10/90% output.
Range -8 dBm to -15 dBm, Wavelength1270-1560 nm
Wavelength 1300 to 1560 nm, Conversion Gain >300
volts/watt, Frequency Response < 3 dB down at 1 GHz.
Wavelength 1200 - 1600 nm, Range 0 - 30 dB
Unique
Unique
Unique
64 kb/s ternary clock.
0 to 200 MHz 3, 6, 10, 20 dB fixed attenuator pads
75Ω coaxial cable 6dB loss at 1 MHz
75Ω coaxial cable 6dB loss at 4 MHz
75Ω coaxial cable 12dB loss at 17 MHz
75Ω coaxial cable 12dB loss at 70 MHz
Recommended
Model
HP 54520A opt 001
or HP 54810A opt 001
HP 3335A opt 001 or
HP 3325B & HP
8657A with 50/75Ω
matching pad
HP 8657A
HP 5325A opt 010
or HP 53181A opt 001
HP 8560E
HP 15477C
HP 8153A and HP
81536A
HP 11982A opt 012
HP 8157A or
HP 8156A opt 100
HP 81000FI (Qty 4)
HP 11871A (Qty 2)
HP 37718A/37719A or
HP 37717C opt UKJ
HP 37732A
HP 86213A
8120-0049 (120 m)*
8120-0049 (80 m)*
8120-0049 (120 m)*
8120-0049 (80 m)*