Pulse Width Performance Test; Equipment - HP 8112A Operating, Programming And Servicing Manual

50 mhz programmable pulse generator
Table of Contents

Advertisement

Max Jitter
0.2% of programmed value ± lOOps

Equipment

Attenuator 20 dB, 2 W.
Test Setup 1
2
Test Se1up
H P 8112A
1 . Set up the HP 81 12A as follows:
Procedure
2 . Connect the HP 8 1 12A and oscilloscope as shown in Figure 8-5 .
3 . Verify that for the following HP81 12A WID settings the scope
4. Record your results on a copy of the Test Record, specifi e d limits
(
)
Counter
HP 5335A
Cable Assembly BNC
Digitizing Oscilloscope
50 !lFeedthrough Termination
(
Required if counter input impedance
HP 8112A
8·5.
Figure
0
0
0
8-6.
Figure
Input Mode
Control Mode
Transition
DEL
HIL
LOL
display indicates that the HP 8 1 12A output is within the specified
limits.
are given here and on the Test Record.
(
X )
2
(
)
HP 54121T
(
2 x )
-f;
50 fl).
Oscilloscope (HP 5412xTI
20 dB Attenuator
with BNC/SMA Adaptor

Pulse Width Performance Test

Counter (H P 5 33 5A)
Pulse Width Performance Test
NORM
Off
Fixed
65 ns
1 .00 v
- 1 .00 v
Testing Performance
8-9