Changing the ROM
Change 42
3. For each ROM in Table B-6 connect the start and stop probes
to the indicated test point and use the data probe to check the
signature at the listed pins :
U9
ROM2
ROM:
Start/Stop:
TP2
Pin
9
OA6P
10
UPlC
1 1
95FH
13
4102
14
AHUA
15
06AF
16
7000
17
4C7P
If the ROMs are changed, the data saved in the RAM has to be
made compatible with the new ROMs. This can be done by setting
the HP 8 1 12A to RCL 0 and turning the instrument off and on
again.
If the instrument becomes totally inoperable switch it off and
disconnect the RAM back-up battery for at least 30 seconds. This
will destroy the stored RAM data. Re-connect the RAM back-up
battery and switch the instrument on. The Standard Parameter Set
is now loaded into the RAM.
A change in manufacturing method introduced a part number change
for a widely used 0.47 µF capacitor.
Table B-6. ROM signatures
US
U7
ROM3
ROM4 U6 ROM5 U5 ROM6
TP3
TP4
U88F
5808
70CU
60F5
6PC3
452P
3C35
7733
U623
41PU
A4HF
3HOH
2C09
P912
2125
8PF7
TP5
TP6
3HA1
P191
3HFH
61A8
03A9
OlUP
34PH
H862
C3FO
61Cl
F8H8
73UF
FP6U
5U29
A795
P3F3
Backdating B-29