HP 8112A Operating, Programming And Servicing Manual page 88

50 mhz programmable pulse generator
Table of Contents

Advertisement

HP-I B Universal
Commands
Note
Programming
6 · 1 8
The HP 8 1 12A supports the following HP-IB Universal commands:
These are HP-IB commands, NOT instrument programming
commands. They are not used in programming messages. If you
require more information on the HP-IB protocol and hardware refer
to "General" for a list of references.
HP-IB
Mnemonic
DCL
SDC
LLO
GTL
GET
UNL
UNT
SPE
SPD
MLA
MTA
An HP-IB D CL command causes the HP 8 1 12A to load its standard
DCL
parameter set. The instrument remains in its current mode (local or
remote).
An HP-IB SDC command causes the HP 8 1 12A to load its standard
SOC
parameter set and enter remote mode.
An HP-IB GET command simulates an external trigger to the HP
GET
8 1 12A in TRIG, E.BUR and E.SWP modes.
Description
Device Clear
Selected Device Clear
Local Lockout
Go to Local
Group Execute Trigger TRIGGER 712 / TRIGGER 7
Unlisten
Untalk
Serial Poll Enable
Serial Poll Disable
My listen address
My talk address
BASIC
5.0/5.1 equivalent
CLEAR 7
CLEAR 712
LOCAL LOCKOUT 7
LOCAL 712 / LOCAL 7
SEND 712;UNL
SPOLL ( 712 )
SEND 712;UNT
selectable
selectable

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents