9-14 Adjustment Procedures
14. On the oscilloscope press (Autoscale ) and adjust A2R8 for
10 ns ± 0.2 ns at 503 of amplitude. See Figure 9-5.
15. Alter HP 8 1 12A WID to 99.9 ns.
16. Adjust the low range end with A2R59 for 101 ns ±0.2 ns.
1 7. Repeat previous six steps and re-adjust if necessary.
18. Alter HP 8 1 1 2A WID to 50 ns.
19. Check that Min. width is 50 ns ± 2ns.