MODEL 8112A
ERRATA (Cont.)
page 8-12, Testing Performance
Output Levels Performance Test change to read:
step 3. Set up the DVM as follows
Function
Add or
Change:
Measurement
HP 3458A setup for peak-voltage measurement:
Function
Trig
MENU ->
MENU
NPLC
->
HP 3456A setup for peak-voltage measurement:
Function
TRIGGER EXT
N CYC INT
page 9-2, Adjustment Procedures
Table 9-1 , Adjustment Procedures - Changeable Components change to read:
Procedure
Overshoot/Transition Times
1 0. July 1 996
DCV
Peak Voltage
DCV
EXT
(press Trig button, select EXT, terminate
with ENTER key)
0.1
(press NPLC button, press period ".",
then " I " keys, terminate with ENTER key)
= V
0. 1
(press period ". ", then " l " keys,
press STORE button,
press CHS (N CYC INT) key)
Reference
Range
A l C532
0 pF (open) to 33 pF
Description
Increasing V532 dereases
transition times and
increases overshoot in
FIXED and LINEA mode
Page 7 of 1 6