Advertisement

Combined operation
5.4 Application planning
5.4.2

Notes on applications planning

Note
Usability of the device
Before you use the device, use the technical specifications to check that your device is
suitable for the measuring task.
Note
Vibrations at the installation site disturb the measured signal
When you use OXYMAT 7 and/or ULTRAMAT 7 analyzer modules, vibrations, especially in
the frequency range from 6 Hz to 15 Hz, can influence the measured signal.
• Select an installation location which is vibration-free.
• If necessary, change the measuring frequency settings:
– Make sure that the measuring frequencies do not match the fundamental of the
– You must also ensure that the measuring frequency differs from half the basic
Note
Magnetic radiance of the OXYMAT 7
Due to its functional principles, the OXYMAT 7 emits magnetic stray fields. Therefore, do not
operate magnetically sensitive devices in the immediate vicinity of the analyzer module.
Depending on the sensitivity, a distance of up to 50 cm from the analyzer module is required.
5.4.3

Operation with at least one OXYMAT 7

Note
Additional clamping screws
Additionally required clamping screws are not included in the scope of delivery. Order the
clamping screws separately.
42
installation site nor one of its harmonic multiples.
frequency of the vibration.
Operating Instructions, 05/2018, A5E31930403-06
Wall-mounted device