Technical specifications ULTRAMAT 23
Enclosure
Weight
Dimensions (W x D x H)
19" rack unit
Bench-top unit
Electrical characteristics
Power consumption
Analyzers in hazardous area
Fuse values
100 to 120 V
200 to 240 V
Climatic conditions (if deviating from general data)
Permissible ambient temperature
During operation
Hydrogen sulfide sensor
During transportation and storage
All analyzer parts/ sensors
except
Hydrogen sulfide sensor
Maximum operating altitude
Pollution degree
Gas inlet conditions
Sample gas pressure
Without pump
With pump
Sample gas flow
Sample gas temperature
Sample gas humidity
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779144002-01, 09/2018
Approx. 10 kg The weight varies according to the ordered variant.
483 mm x 378 mm x 177 mm
529 mm x 342 mm x 170 mm
Max. 60 VA depending on ordered version
It has to be ensured that the transient protection is set to a value that
does not exceed 140 % of the measured peak voltage value at the de-
vice power supply connectors.
1.25T/250
0.63T/250
+ 5 ... +40 °C
-20 ... +60 °C
-10 ... +55 °C
2000 m above sea level
2
Unpressurized (< 1200 hPa absolute), gas outlet must be unpressurized
Unpressurized suction mode, set in factory with 2 m (6 1/2 ft) hose
at sample gas outlet; full-scale value calibration necessary under
different venting conditions
72 ... 120 l/h (1.2 .. 2 l/min)
0 ... 50 °C (32 ... 122 °F)
< 90% RH (relative humidity), non-condensing
27