7.2
Technical specifications ULTRAMAT 23
Enclosure
Weight
Dimensions (W x D x H)
19" rack unit
•
Bench-top unit
•
Electrical characteristics
Power consumption
Analyzers in hazardous area
Fuse values
100 to 120 V
•
200 to 240 V
•
Climatic conditions (if deviating from general data)
Permissible ambient temperature
During operation
•
–
Hydrogen sulfide sensor
During transportation and storage
•
–
All analyzer parts/ sensors
except
–
Hydrogen sulfide sensor
Maximum operating altitude
Pollution degree
Gas inlet conditions
Sample gas pressure
Without pump
•
With pump
•
Sample gas flow
Sample gas temperature
Sample gas humidity
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779144002-01, 09/2018
Approx. 10 kg The weight varies according to the ordered variant.
483 mm x 378 mm x 177 mm
529 mm x 342 mm x 170 mm
Max. 60 VA depending on ordered version
It has to be ensured that the transient protection is set to a value that
does not exceed 140 % of the measured peak voltage value at the de-
vice power supply connectors.
1.25T/250
0.63T/250
+ 5 ... +40 °C
-20 ... +60 °C
-10 ... +55 °C
2000 m above sea level
2
Unpressurized (< 1200 hPa absolute), gas outlet must be unpressurized
Unpressurized suction mode, set in factory with 2 m (6 1/2 ft) hose
at sample gas outlet; full-scale value calibration necessary under
different venting conditions
72 ... 120 l/h (1.2 .. 2 l/min)
0 ... 50 °C (32 ... 122 °F)
< 90% RH (relative humidity), non-condensing
27