Installation Location Requirements - Siemens 6 Series Compact Operating Instructions

Table of Contents

Advertisement

3.2

Installation location requirements

CAUTION
Direct sunlight
Device damage.
The device can overheat or materials become brittle due to UV exposure.
Protect the device from direct sunlight.
Make sure that the maximum permissible ambient temperature is not exceeded. Refer to the information in Technical
specifications (Page 26).
WARNING
Insufficient air circulation
Risk of fire.
The device may overheat or start burning in case of insufficient air circulation.
Ensure sufficient air circulation in the room. Refer to the information in Technical specifications (Page 26).
CAUTION
Strong vibrations
Vibrations can influence the result of a measurement.
The analyzer must therefore only be installed at a location which is free of vibration.
Ensure that there is little vibration when using the device. The maximum vibrations must not exceed 5 m/s².
DANGER
Explosion hazard
If a flammable or ignitable atmosphere exists, plugs must never be disconnected or lamps/fuses replaced when the
analyzer is supplied with power.
ATEX Zone 2
The following applies to devices which are operated according to ATEX in Ex zone 2:
DANGER
Explosion hazard
The gas analyzers for use in Ex zone 2 must be installed in a lockable enclosure. This enclosure must comply with the
requirements of EN 60079-7 and must be designed for all ambient conditions which can occur during operation. This
enclosure only be opened using a tool (e.g. a key).
The ambient temperature is
5 °C to 55 °C for the CALOMAT 6E with article number 7MB25x-xGAxx
5 °C to 45 °C for all other devices.
Suitable measures must additionally be applied to ensure that
The generation of potentially explosive gas mixtures inside the analyzer does not exceed the specifications of Zone 2.
Interferences cannot lead to a deviation of more than 40 % from the rated voltage.
The location of the device has a maximum degree of pollution 2.
The maximum permissible overvoltage category II is adhered to.
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779144002-01, 09/2018
9