Siemens 6 Series Compact Operating Instructions page 32

Table of Contents

Advertisement

Parts in gas path wetted by sample gas
Gas connections
Gaskets (O–rings)
Sensor
Measuring cell body
Internal gas path (gas lines)
Leak-tightness
Gas inlet conditions
Permissible sample gas pressure
Sample gas flow
Sample gas temperature
Sample gas humidity
7.7
Technical specifications CALOMAT 62E
Enclosure
Weight
Dimensions
Electrical characteristics
Power consumption (unit)
Fuse values
100 ... 120 V
200 ... 240 V
Gas path
Gas connections
Gaskets (O–rings)
Sensor
Internal gas path (gas lines)
Leak-tightness
Gas inlet conditions
Sample gas pressure
Sample gas flow
Sample gas temperature
Sample gas moisture content
Time response
Warm-up time
Delayed display T
(incl. dead time)
90
Damping (electronic time constant)
Dead time (at 1 l/min)
32
Stainless steel 1.4571, pipe diameter 1/4" or 6 mm
FFKM (e.g. Chemraz)
Silicon, SiO
N
, gold, epoxy resin, glass
x
y
Stainless steel 1.4571
Stainless steel 1.4571
Loss through leaks < 0.5 μl/s
800 to 1100 hPa absolute
30 to 90 l/h (0.5 to 1.5 l/min)
0 to 50 °C
< 90% RH (relative humidity), non-condensing
Approx. 13 kg
383 x 177 x 390 mm
Approx. 30 VA
T1L250V
T630L250V
Female thread 1/8"-27 NPT (according to ANSI B 1.20.1)
FKM (e.g. Viton) or FFKM (e.g. Kalrez)
Glass
Stainless steel 1.4571
Loss through leaks < 0.5 μl/s
800...1100 hPa (absolute)
30...90 l/h (0.5...1.5 l/min)
0 to 50 °C
< 90% RH
< 30 min, maximum accuracy is achieved after about 2 hours
Approx. 35 s
0 to 100 s, adjustable
Approx. 34 s (because of diffusion to sensor)
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779144002-01, 09/2018

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents