Leadwire Installation; Electrode Attachment - GE ApexPro Operator's Manual

Telemetry system
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Leadwire installation

Electrode attachment

2001989-200E
CAUTION
FALSE ALARMS—If clinical workflow includes pre-admitting
telemetry transmitters, attach ECG leadwires to patient prior to
attaching ECG leadwires to transmitter. Failure to follow this
sequence may result in false alarms or interruption in ECG
waveform.
The transmitter can use the following Multi-Link leadwire sets:
Multi-Link 6-leadwire set
Multi-Link 5-leadwire set
Multi-Link 3-leadwire set
To install a leadwire set into the transmitter, align the leadwire pins with the
connector on the top of the transmitter, then push the leadwire set firmly into the
transmitter.
To disconnect the leadwire set from the transmitter, grasp the molded end or the
combiner firmly and pull away from the transmitter.
1.
Attach leadwires to the transmitter by plugging the Multi-Link leadwire set into
the transmitter.
To use sets of Multi-Link individual leadwires, firmly press the individual
leadwires into their appropriate locations on the combiner. Use the colors on the
leadwires to place them in corresponding order with the colors that appear on the
back of the transmitter.
ApexPro™
Equipment setup
307C
308B
3-5