System setup
4-16
Option
Unit License Default: Full
Disclosure License Type
Offline Storage
Start Data Storage
Bed List
Restore
Current telemetry listings
NOTE
For important configuration information, refer to the appropriate CIC Pro center
service manual.
NOTE
Telemetry beds are distinguished from monitoring beds by an asterisk appended
to the end of the bed number.
To view the Current Telemetry Listings, click CIC Setup > Current Telemetry
Listings.
ApexPro™
Function
NOTE
You must be in the Service mode at the CIC Pro center
to modify this setting.
Display a list of the full disclosure license type. Choices are:
none, 24 hours, 48 hours, and 72 hours.
NOTE
If the default does not match the actual license, full
disclosure does not work.
Select a time period to store full disclosure data if contact
with a monitor has been lost. Choices are: 5 mins, 30 mins,
1, 2, 4, 8 and 12 hours.
WARNING
POTENTIAL DATA LOSS — Do not allow a NO COMM
(patient offline) event to exceed the time limit selected in the
Offline Storage setting. The patient's full disclosure data is
deleted if the time limit is exceeded.
For more information, refer to the appropriate CIC Pro
center service manual.
NOTE
You must be in the Service mode at the CIC Pro center
to modify this setting.
Designate how full disclosure is enabled for patients at the
time of admission. Choices are: automatically for all beds,
automatically if listed, and manually.
NOTE
You must be in the Service mode at the CIC Pro center
to modify this setting.
Lists beds for which full disclosure data is automatically
stored.
Clear any changes you made to the full disclosure default
settings and revert to the previous settings.