Technical data
9.1 Device technical data
Construction
Weight (without options)
Dimensions (W x H x D) in mm (inches)
Housing material
Electrical connection via terminal screws
Cable inlet via plastic cable glands
Power supply
Supply
Rated voltage
Operating range
Current consumption
For connection to approved intrinsically safe electrical circuits with:
We recommend Turck MK72-S13-EXO/24VDC solenoid drivers
Internet address www.turck.com
Certificates and approvals
Protection against explosion in accordance with EN50014, EN50020, EN50021
Intrinsic safety "i"
Identification
Permissible ambient temperature
80
Maximum supply voltage U
Maximum short circuit current I
Maximum power P
i
Internal capacity C
i
Internal inductance L
i
Temperature class T4, T5
Temperature class T6
approx. 2.5 kg
172 x 320 x 80 (6.8 x 12.6 x 3.2)
Macrolon (polycarbonate + 20 % glass fiber)
surface coated with CrNi layer and varnished
Rigid 2.5 mm (0.984 inch)
Flexible 1.5 mm (0.590 inch)
Flexible with connector sleeves 1.5 mm
(0.590 inch)
1 x Pg 13.5 for 1 cable, ∅ 6 ... 12 mm
(0.236 ... 0.472'') (e.g. PROFIBUS)
1 x Pg 13.5 for 3 cables, ∅ 4 mm (0.157
inches)
1 x Pg 11 for 1 cable, ∅ 4 ... 10 mm
(0.157 ... 0.394 inches)
4 x Pg 11 for 2 cables each, ∅ 4 mm (0.157
inches) (e.g. sensors)
Clamps 11 and 12
DC 16 V
DC 15 ... 17 V
< 40 mA
≤ 17.4 V
i
≤ 191 mA
i
≤ 1.35 W
≤ 33 nF
≤ 28 µH
TÜV 06 ATEX 2952
II 2(1) G EEx ia [ia] IIC T4/T5/T6
-20 °C ... +60 °C (-4 °F ... 140 °F)
-20 °C ... +50 °C (-4 °F ... 122 °F)
SITRANS DA400 for material flow monitoring with protection against explosion