B30 Patient Monitor
To separate the LCD display from
the frame unit
6-14
.
Document no
2044677-001
4. Place the monitor face down on a non-adrasive, static-free
surface.
NOTE: Be careful not to damage the ComWheel.
5. Lift the back cover up.
1. Remove the three (T8) screws from the borders of the frame
unit, one on the other, two on the other side of the frame).
2. Separate the frame assembly from the front cover unit.
NOTE: The backlight circuit runs on high voltage. Do not touch
the inverter board or the backlight tube leads when powered.
3. Disconnect the displayed cable.
4. Remove the four (T10) screws holding the display to the
frame unit.
5. Lift the display carefully up.
NOTE: When reassembling the LCD display, be careful that no
dirt or finger prints are left between the LCD display element
and the protection glass window.
NOTE: Do not use excessive force when fastening the display to
the frame. Fastening the screws too tightly might bend the
display too much and break it. The screws must also be
fastened gradually, first fasten all screws slightly, then more
tightly.