B30 Patient Monitor
2
Specifications
2.1 General specifications
N-FCREC
Module size, W × D × H
Module weight
Power consumption
N-FREC
Module size, W × D × H
Module weight
Power consumption
N-FC
Module size, W × D × H
Module weight
Power consumption
2.1.1 Environmental specifications
Operating temperature
Storage temperature
Atmospheric pressure
Humidity
Protection against electrical shock
2.1.2 Functional alarms
Funtional alarms for
−
Blocked sample line
−
D-Fend replacement
−
D-Fend check
2.2 CO
measurement
2
2.2.1 Typical performance
EtCO
2
FiCO
2
Measurement range
Accuracy
CO
concentration 0 to 15 vol%
2
CO
concentration 15 to 20 vol%
2
Measurement rise time
Adjustable low and high limits for EtCO
8-2
.
Document no
2044677-001
90 × 169 × 191 mm, 3.54 × 6.65 × 7.51 in
1.0 kg/2.2 lb.
2.7 W standby, 11.5 W printing
90 × 169 × 191 mm, 3.54 × 6.65 × 7.51 in
0.8 kg/1.7 lb.
1.2 W standby, 10 W printing
46 × 169 × 191 mm, 1.81 × 6.65 × 7.51 in
0.5 kg/1.1 lb.
1.5 W typical
+10...+40 °C (+50 to 104 °F)
-20...+60 °C (-4 to +140 °F)
666...1060 hPa /
(67...106 kPa)
(500...800 mmHg)
(666...1060 mbar)
10...95% non-condensing (in airway 0...100%, condensing)
Type BF
End-tidal CO
concentration
2
Inspired CO
concentration
2
0 to 20 vol% (0 to 20 kPa, 0 to 150 mmHg)
±(0.2 vol% +2% of reading)
±(0.7 vol% +2% of reading)
< 300 ms with nominal flow
and FiCO
.
2
2