Functional Check; Visual Inspection; Functional Inspection; General - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
3

Functional check

These instructions include procedures for a functional check for B30 monitor. The functional
check is recommended to be performed after monitor installation.
If these check haven't resolved your problem, please refer to the
more detials
These instructions include a
procedures.
An electrical safety check and a leakage current test are recommended to be performed prior
to the monitor installation.

3.1 Visual inspection

Make sure that the monitor is switched to standby.
Disconnect the mains power cord from the monitor.
If the monitor is connected to the Network, disconnect the RJ-45 cable from the monitor.
1.
Check all units visually
Check that all parts are intact and that the cables and screws are connected and
tightened properly. Especially check the following parts:
Check that modules go in smoothly and lock up properly.

3.2 Functional inspection

3.2.1 General

1.
Connect the mains power cord to the monitor.
Check that the Mains power LED is lit.
2.
Switch the monitor on.
Check that the monitor starts up properly, i.e. a normal start-up sound is heard from the
loudspeaker, the alarm LEDs turn on and off, and the monitoring screen appears. No error
messages should appear on the screen.
3.
Configure the screen for the parameters that are connected.
4.
Enter the Service Menu.
When applicable, check from the corresponding Parameters submenu that the Timeouts,
Bad checksums and Bad c-s by mod values of inserted modules are not increasing faster
than by 5 per second. Check also that the module memories have passed the internal
memory test, i.e. RAM, ROM and EEPROM all state OK.
If connected, the recorder should record two lines of start-up information.
Preset the measurement settings for those parameters that are connected, for example:
Print/Recorder
3-14
.
Document no
2044677-001
"Functional check form,
sampling line is connected to the extension module.
- Record Waveforms - Waveform 1 - ECG1
"Planned
maintenance" for
B30" to be filled in when performing the