Electromagnetic Emissions And Immunity; Emc Guidance - Philips DL8740/15 Manual

Table of Contents

Advertisement

Indicates the manufacturer, as defined in
EU Directives 93/42/EEC.
Symbol for the 'Bluetooth Combination
mark'. The device uses Bluetooth for
communication.
Indicates the manufacturer's serial
number so that a specific medical device
can be identified.
Symbol for indoor use only.
Indicates caution.The user should consult
the instructions for use for important
cautionary information such as warnings
and precautions that cannot, for a variety
of reasons, be presented on the medical
device itself.
IP22: The first number 2: Protected against
solid foreign objects of 12,5 mm Ф and
greater. The second number: Protected
against vertically falling water drops when
enclosure is tilted up to 15°. Vertically
falling drops shall have no harmful effects
when the enclosure is tilted at any angle
up to 15° on either side of the vertical.
Indicates the storage and transportation
60ºC
temperature limits to which the medical
–20ºC
device can be safely exposed: -20°C to
60°C.
Indicates the storage and transportation
relative humidity limits to which the
medical device can be safely exposed:
15% - 85%
Symbol for the 2 year Philips guarantee.
The Green Dot ('Der Grüne Punkt' in
German) is the license symbol of a
European network of industry-funded
systems for recycling the packaging
materials of consumer goods.
Electromagnetic emissions and
immunity
The device is approved according to EMC safety
standard EN 60601-1-2. It is designed to be used in
typical domestic environments.
Hereby, Philips Consumer Lifestyle BV declares
that the radio equipment type Bluetooth LE is in
compliance with Directive 2014/53/EU.
The full text of the EU declaration of conformity is
available at the following internet
address: www.philips.com/support

EMC Guidance

- The ear thermometer needs special precautions
regarding EMC and needs to be installed and
put into service according to the EMC
information provided in this document.
- Wireless communications equipment such as
wireless home network devices, mobile phones,
cordless telephones and their base stations,
walkie-talkies can affect this equipment and
should be kept at least a distance d = 3.3 m
away from the equipment.
Note: As indicated in IEC 60601-1-2:2007 for ME
equipment, a typical cell phone with a maximum
output power of 2 W yields d = 3.3 m at an
immunity level of 3V/m.
Table 1 Guidance and manufacturer's
declaration – electromagnetic emissions
- for all ME equipment and ME systems
Guidance and manufacturer's declaration –
electromagnetic emissions
The device is intended for use in the
electromagnetic environment specified below. The
customer or the user of the device should assure
that it is used in such an environment.
Emissions test Com­
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions IEC
61000-3-2
Voltage
fluctua­
tions/flicker
emissions IEC
61000-3-3
Table 4 Guidance and manufacturer's
declaration – electromagnetic immunity
–for ME equipment and ME systems that
are not life supporting
Guidance and manufacturer's declaration –
electromagnetic immunity .The device is intended
for use in the electromagnetic environment
specified below. The customer or the user of the
device should assure that it is used in such an
environment.
IMMUNITY
test
Conducted RF
IEC
61000-4-6
Radiated RF
IEC
61000-4-3
Electromagnetic environment - guidance
pli­
ance
Group 1 The DL8740 uses RF
Class B
Not
applica
ble
Not
applica
ble
IEC 60601
TEST LEVEL
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to 2.5
GHz
Electromagnetic
environment -
guidance
energy only for its
internal function.
Therefore, its RF
emissions are very
low and are not likely
to cause any
interference in nearby
electronic equipment.
Compliance
level
3 Vrms
3 V/m

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dl8740

Table of Contents