Advertisement

Problem
 
 
 
The battery
symbol flash
es on the dis
play.
The display
shows Err0 
The display
shows Err2
The display
shows Err3
 

Explanation of symbols

The warning signs and symbols are essential to
ensure that you use this product safely and correct
ly and to protect you and others from injury. Below
you find the meaning of the warning signs and sym
bols on the label and in the user manual.
Symbol for 'follow instructions for use'.
This symbol means that the part of the
device that comes into physical contact
with the user (also known as the applied
part) is of type BF (Body Floating) accord
ing to IEC 60601-1.
Symbol for 'the device complies with Euro
pean Medical Device Directive 93/42/EEC
requirements'. 0344 refers to the notified
body.
Symbol for WEEE, waste electrical and
electronical equipment. Electrical waste
products should not be disposed of with
household waste. Please recycle where
facilities exist. Check with your local
authority or retailer for recycling advice
and see chapter 'Battery recycling'.
This symbol means that this product con
tains batteries which shall not be disposed
of with normal household waste
(2006/66/EC).
Indicates the manufacturer, as defined in
EU Directives 93/42/EEC.
Symbol for the 'Bluetooth Combination
mark'. The device uses Bluetooth for com
munication.
Indicates the manufacturer's serial num
ber so that a specific medical device can
be identified.
Symbol for indoor use only.
Indicates caution.The user should consult
the instructions for use for important cau
tionary information such as warnings and
precautions that cannot, for a variety of
reasons, be presented on the medical
device itself.
This symbol on the device means: protect
ed against access to hazardous parts with
a finger and against vertically falling water
drops when tilted up to 15 degrees.
Indicates the storage and transportation
60ºC
temperature limits to which the medical
–20ºC
device can be safely exposed: -20°C to
60°C.
Symbol for the 2 year Philips guarantee.
The Green Dot ('Der Grüne Punkt' in Ger
man) is the license symbol of a European
network of industry-funded systems for
recycling the packaging materials of con
sumer goods.
Electromagnetic emissions and
immunity
The device is approved according to EMC safety
standard EN 60601-1-2. It is designed to be used in
typical domestic environments.

EMC Guidance

- The ear thermometer needs special precautions
regarding EMC and needs to be installed and
put into service according to the EMC informa
tion provided in this document.
- Wireless communications equipment such as
wireless home network devices, mobile phones,
cordless telephones and their base stations,
walkie-talkies can affect this equipment and
should be kept at least a distance d = 3.3 m
away from the equipment.
Note: As indicated in IEC 60601-1-2:2007 for ME
equipment, a typical cell phone with a maximum
Possible
cause
Your room
temperature
is too low or
too high.
You are using
the device
outdoors.
You have
held the
device in
your hand
too long.
The batteries
are empty.
 Self-test
error
Stabilization
error
Battery level
low
Solution
Use your ther
mometer at room
temperatures
between 10.0°C
and 40.0°C
(50.0°F and
104.0°F).
The device is only
intended for
indoor use.
Put the device on
the table in the
room where the
measurement is
taking place and
let it cool down
first.
Replace the emp
ty batteries with
new ones.
Remove and
replace the batter
ies. If the error per
sists, contact
Philips Consumer
Care Center in
your country .
Wait 30 minutes
and measure
again.
Replace the emp
ty batteries with
new ones.