Kenwood CS-1022 Instruction Manual page 12

Hide thumbs Also See for CS-1022:
Table of Contents

Advertisement

C S - 1 0 2 2 , 1 0 2 1 , 1 0 1 2
(§)
COUPLING
Three-position lever switch; selects coupling for sync
trigger signal.
A C :
Trigger is ac coupled. Blocks dc component
of input signal; most commonly used posi-
tion.
VIDEO
FRAME:
Vertical sync pulses of a composite video
signal are selected for triggering.
VIDEO
LINE:
Horizontal sync pulses of a composite video
signal are selected for triggering. The LINE
position is also used for all non-video
waveforms.
(§)
SOURCE
Five-position lever switch; selects triggering source
for the sweep, with following positions;
V. MODE: The trigger source is determined by ver-
tical MODE selection.
C H 1 :
CH2:
ADD:
12
CS-1022, 1021, 1012
Channel 1 signal is used as a trig-
ger source.
Channel 2 signal is used as a trig-
ger source.
The algebraic sum of channel 1
and channel 2 signal is the trigger
source. (If CH2 INV engaged, the
difference
becomes the trigger
source.)
Fig. 5
C S - 1 0 2 0 , 1 0 1 0
@)
COUPLING
Three-position lever switch; selects coupling for sync
trigger signal.
A C :
Trigger is ac coupled. Blocks dc component
of input signal; most commonly used posi-
tion.
VIDEO
FRAME:
Vertical sync pulses of a composite video
signal are selected for triggering.
VIDEO
LINE:
Horizontal sync pulses of a composite video
signal are selected for triggering. The LINE
position is also used for all non-video
waveforms.
(§)
SOURCE
Three-position lever switch; selects triggering source
for the sweep, with following positions;
INT:
Waveform being observed is used as sync
trigger.
LINE:
S w e e p
( 5 0 / 6 0 Hz).
EXT:
Sweep is triggered by signal applied to EXT
TRIG INPUT jack ® .
is
t r i g g e r e d
by
line
v o l t a g e

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cs-1021Cs-1012Cs-1020Cs-1010

Table of Contents