Kenwood CS-1022 Instruction Manual page 7

Hide thumbs Also See for CS-1022:
Table of Contents

Advertisement

ditions could also pose a safety hazard, which the
ground cable will prevent.
5. Always use the probe ground clips for best results. Do
not use an external ground wire in lieu of the probe
ground clips, as undesired signals may be introduced.
6. Operation adjacent t o equipment which produces strong
ac magnetic fields should be avoided where possible.
This includes such devices as large power supplies,
transformers, electric motors, etc., that are often found
in an industrial environment. Strong magnetic shields
can exceed the practical CRT magnetic shielding limits
and result interference and distortion.
Plug configuration
Power cord and plug type
North American
120 volt/60 Hz
Rated 15 amp
(12 amp max; NEC)
Universal Europe
220 volt/50 Hz
Rated 16 amp
U.K.
240 volt/50 Hz
Rated 13 amp
Australian
240 volt/50 Hz
Rated 10 amp
North American
240 volt/60 Hz
Rated 15 amp
(12 amp max; NEC)
Switzerland
240 volt/50 Hz
Rated 10 amp
Fig. 1 Power Input Voltage Configuration
7. Probe compensation adjustment matches the probe t o
the input of the scope. For best results, compensation
of probe should be adjusted initially, then the same pro-
be always used w i t h the input of scope. Probe compen-
sation should be readjusted whenever a probe from a
different scope is used. (See page 19)
8. Do not depress t w o or more push-switches of the ver-
tical MODE switch simultaneousely. (Except ADD)
Damage
t o
i n s t r u m e n t
C S - 1 0 2 1 , CS-1012)
9. In X-Y operation, do not pull out the PULLx 10 MAG
s w i t c h . If pulled out it, noise may appeare on the
waveform.
Factory installed
instrument fuse
0.8 A, 250 V
Fast blow
AGC/3AG
0.5 A, 250 V
Fast blow
5 x 20 mm
0.5 A, 250 V
Fast blow
5 x 20 mm
0.5 A, 250 V
Fast blow
5 x 20 mm
0.5 A, 250 V
Fast blow
AGC/3AG
0.5 A, 250 V
Fast blow
AGC/3AG
5 x 20 mm
may
result.
( C S - 1 0 2 2 ,
Line cord plug fuse
None
None
0.5 A
TypeC
None
None
None