HP 16532A Service Manual page 30

1 gsa/s digitizing oscilloscope
Table of Contents

Advertisement

Procedure:
1. Using a BNC adapter, connect one end of the cable to the power supply. Connect the BNC tee,
the blocking capacitor, and the shorting endcap to the other end of the cable. Refer to figure 3-1.
2. In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set
the following parameters in the order given.
Menu
Channel
Trigger
Display
Set the Channel Coupling field to 1 M Ω /DC or damage to the equipment will
result.
Figure 3-1. Voltage Measurement Accuracy Test
Selection
Input
Probe
Coupling
s/Div
Mode
Mode
Average #
Grid
Markers
HP 16532A - Performance Tests
Setting
1
1:1
1 MΩ / DC
500 ns
Immediate
Average
8
On
On
3-7