HP 16532A Service Manual page 63

1 gsa/s digitizing oscilloscope
Table of Contents

Advertisement

3. Install the Composite PV Test software disk in the inactive disk drive.
4. Touch the box labeled Touch box to Load Test System. See figure 6-3.
When the self-test software is loaded, the user operating system is overwritten in
the mainframe system RAM.
5. From the test system screen in figure 6-4, touch Test System, then touch 1 GSa/s SCOPE D. (For
the example shown in figure 6-4, the oscilloscope module is in slot D. If multiple oscilloscope mod-
ules are present, select the one to be tested.)
Note
Figure 6-3. Load Test System Screen
Figure 6-4. Test System Screen
HP 16532A - Service
6-5