Additional adjustments -
Date/Time
Additional adjustments
- Reset
Setup
Menu item
Lock adjustment
Measurement loop name
Units
Probe length
Type of medium
Application
Medium, dielectric constant
Superimposed gas phase
PBD-51041057
In this menu item, the internal clock of the sensor is adjusted.
With a reset, certain parameter adjustments carried out by the user
are reset.
The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of ship-
ment from the factory incl. the order-specific settings. A created false
signal suppression, user-programmable linearization curve as well as
the measured value memory will be deleted.
Basic settings: Resetting of the parameter settings, incl. special
parameters, to the default values of the respective instrument. Any
stored false signal suppression or user programmable linearisation
curve, as well as the measured value memory, is deleted.
The following table shows the default values of the instrument. De-
pending on the instrument version or application, all menu items may
not be available or some may be differently assigned:
Default value
Released
Sensor
Distance unit: mm
Temperature unit: °C
Length of the probe Ex factory
Bulk solid
Level in the metallic vessel
Granules, powder, cement /
1.5 ... 3
Yes
SITRANS LG260 - Operating Instructions
Modified value
39