Siemens SITRANS LG260 Operating Instructions Manual page 71

Guided wave radar
Hide thumbs Also See for SITRANS LG260:
Table of Contents

Advertisement

15 mm
(0.591
"
)
2 mm
(0.079
")
0
-2 mm
(-0.079
")
1
-10 mm
(-0.394
")
0,08 m
(3.15
)
"
Fig. 45: Deviation SITRANS LG260 in rod version
1
Dead band - no measurement possible in this area
L
Probe length
15 mm
(0.591
")
2 mm
(0.079
")
0
-2 mm
(-0.079
")
1
10 mm
(0.394
")
0,08 m
(3.15
")
Fig. 46: Deviation SITRANS LG260 in cable version in water
1
Dead band - no measurement possible in this area
L
Probe length
Repeatability
Variables influencing measurement accuracy
Specifications for the digital measured value
Temperature drift - Digital output
Additional deviation through electromag-
netic interference acc. to EN 61326
Specifications apply also to the current output
Temperature drift - Current output
Also for the additional current output (optional)
7)
PBD-51041057
0,3 m
(11.811
)
"
0,3 m
(11.811
")
≤ ±1 mm
±3 mm/10 K relating to the max. measuring range or
max. 10 mm (0.394 in)
< ±10 mm (< ±0.394 in)
±0.03 %/10 K relating to the 16 mA span max. ±0.3 %
SITRANS LG260 - Operating Instructions
L
L
7)
0,02 m
(0.787
")
0,1 m
(3.94
")
71

Advertisement

Table of Contents
loading