4.10 Establish Communications with GCP

3. Ensure the sample system has been properly pressure tested at this site.
4. Ensure that all SSO (Sample Shut Off) and ARV (Atmospheric Reference Valves) are
working properly.
5. Ensure that all sample system vents and drains are properly oriented.
6. Inspect the sample system to ensure it is in accordance with the sample system
documentation.
7. Circulate and purge process sample through the sample system for a short period of time
before applying to the analyzer. Ensure there is no vapor visible in liquid streams or liquid
visible in vapor streams at the bypass rotameters.
8. Open the valve to the analyzer slipstream and allow sample flow to the analyzer.
4.10
Establish Communications with GCP
1. Refer to the section on assigning IP and network Mask Addresses and input the desired
address into the Maxum II.
2. Launch GCP on the workstation.
3. If analyzer with the IP address entered in the Maxum II does not appear in the analyzer list
at the bottom of the main screen in GCP, add it manually via the menu via,
Unit ;
– a dialog box will appear; enter the unit name, type and IP address,
– then click
4. Launch EZChrom on the workstation.
5. From the main window of EZChrom, create a new instrument from the menu:
– Press
– Right click on the icon and select
– In the dialog box, press the
– In the next dialog box, press the
44
WARNING
Vent all lines and oven exhaust to a location outside the shelter where the pressure is
atmospheric, and in accordance with local and plant operating codes.
Failure to properly vent exhaust gasses may result in ignition or explosion, serious injury
or death. Connection of oven or analyzer vents to a variable pressure source such as a
flare header should not be done as chromatographic performance will be severely
affected. Backpressure on the oven vent could cause the T-rating of the instrument to be
exceeded, which might become an ignition source. In addition, in some situations it may be
possible for backpressure on oven vents to force hazardous gases into the shelter. Refer
to the comments on Hazardous gases earlier in this manual and in the MAXUM II Explosion
Protection Safety Standards Manual
42017542/).
Add and the unit will be added to the analyzer list.
File > New > Instrument . An icon representing an analyzer will appear in the main
screen.
OK .
press
(https://support.industry.siemens.com/cs/document/
Configure > Instrument .
Configure button on the upper right.
Set database button, enter the analyzer IP address, and