Initial Startup Procedures
4.12 Calibrate the Maxum II
Figure 4-5
As each peak is selected on the left pane, the calibration settings for that peak are displayed on
the right. The example below is designed to give the user a basic idea of modifying a calibration
and performing an offline calibration based on the modified settings.
Example Scenario:
A new calibration standard has been installed with slightly different concentrations than the
previous standard. You need to change the calibration levels and recalibrate the current
method.
User Actions:
Perform the following actions to change the calibration levels:
● Select Peaks from the Method on the Navigation Menu and then select the first detector.
This displays peaks for the first detector.
● Click the Calibration checkbox to see calibration settings.
● Select the first peak (in this case Carbon Monoxide).
● In the Calibration Level settings, change the Amount attribute to match the new calibration
standard.
46
Peak Screen - Calibration
Installation Manual
Installation Manual, 12/2019, 2000595 Rev 1