Publication no. 201101
We want to create a COMM field that shall display the FOCUS. Create a new field and under
fielddefinition select Type COMM. The Mammomat® 3000 is sending ether 0 or 1 as the FO-
CUS parameter and the Network ID Camera is translating it to a text according to the proto-
col set-up above. In this case the texts are SF for 0 and LF for 1. With reference to this
settings, the length should be set to 2 and Comm to FOCUS. This means that we have to
define the length of this COMM field to two. Select Comm FOCUS will link the focus data to
this field.
Fielddefinition
Type: Comm
Row:
Comm: Focus
The field+ data do not include the unit (kV, mAs etc.). To add this, a F-text field with the ap-
propriate text has to be inserted after or before the COMM field. Create a new field and un-
der fielddefinition select Type F-text. Enter values for the row and column to place this field
at a position of your preference. Hint, use the Place function to position fields, see chapter
3.4.2 'Picture Win1/Win2'. Enter appropriate text under Fixed text.
Fielddefinition
Type: F-text
Row:
Fixed text:
Focus
These fielddefinitions will give us the text "Focus" followed by the "FOCUS"-value at the cen-
ter bottom of the patient data window.
3.4.2.2 Protocol set-up SIEMENS Mammomat® 300
Define text for Mammomat 300
Small focus: SF
Large focus: LF
This window is used to define the focus text string. The Mammomat® 300 sends this param-
eter as 0 and 1, whereas it should be presented as text instead. This window is used to
define this text. For more detailed information, please refer to the Mammomat® 300 manual.
The string may be up to 8 characters long. Use the up and down arrows to go to the row
below or above and enter appropriate text.
Creating the picture that will be printed on the film is done in the same way as usual selecting
the COMM field type.
October 2005
Length:
56
Column:108
Previous
Next
56
Column: 78
16
NIC2-2 · Installation and Set-up Manual
Field:01
2
Font: Small
Save
Field:02
Font: Small
Save
Save