Niclan Mode - Siemens NIC2-2 Installation And Startup Manual

Network id camera
Hide thumbs Also See for NIC2-2:
Table of Contents

Advertisement

Publication no. 201101
3.5.1

NICLan Mode

NICLan Mode settings
Dicom
NICLan : 014B79
Licence: 0000-0000
Group
Move between the different settings by pressing up- and down arrow keys. Use the Previous
(F2) and Next (F3) keys to toggle between the alternatives. Press F4-key when done.
!
If the Dicom selection is set to "Enable" a valid Licence key MUST be entered.
If not, an error message will appear when trying to leave this window.
To be able to exit this window, make sure correct Licence key value is entered
or set Dicom to "Disabled"
Activates the DICOM protocol. If activaded a valid Licence key must be entered.
DICOM
Unique key generated by the NICLan Module.
NICLan
Licence Licens key for DICOM.
Filter. Name entered here will be compared with "GroupeName" in Download
Group
booking data string sent from the NICLan Module to the NIC2-2. Ref. "TCP/IP
FTP Communication NICLan" specification.
Activates the FTP function.
FTP
Protocol Selects how to handle received patient data. For details, please refer to the Net-
work Installation Manual.
Download =
Direct =
Exp.coll Activates the "exposure save" function. All patient- and exposure-data available
in the ID Camera are stored on the CompactFlash card in the NICLan Module
when a film is marked.
Mem.full Action when out of memory for the "Exp. Coll" function.
Stop =
Overwrite =New exposure data will overwrite already stored data.
!
When DICOM and/or FTP are activated, the HOST channel of the NIC will au-
tomatically be set to communicate with the NICLan Module through the NET
connector. The selection "HOST" under "Setup\Communication param" will
not be selectable.
October 2005
: Disabled
FTP
Protocol :Download
Exp.coll :Disabled
Mem.full :Stop
:
Patient data to the internal bookinglist. Mandatory if DICOM.
Patient data direct to the exposure window.
Not possible to mark a film until stored exposure data is collected.
24
NIC2-2 · Installation and Set-up Manual
:Disabled
Exit