Siemens 30 64 581 B5310 Maintenance Instruction page 21

Table of Contents

Advertisement

Safety inspection
NOTE
NOTE
WARNING
WARNING
If covers were removed, they are to be reattached.
Perform the leakage current measurement according to ARTD-002.731.17... .
The highest value must be entered in the leakage current report.
This value must not exceed the permissible leakage current values according to DIN
VDE 0751-01:2001-10, Table F.1, line "general leakage current information",
of 0.5 mA.
Measure and record the current line voltage. If the measured line voltage value devi-
ates from the line voltage, correct the measured value to the value corresponding to a
measurement at the nominal value of the line voltage. This is also to be documented.
Document the measuring procedure (differential measurement or direct measurement)
and the measuring device used (designation and serial number).
In the case of repeat measurements, the measured value is also to be evaluated.
Siemens AG
Medical Solutions
The leakage current measurement for monitors is to be performed
when these are externally mounted (e.g. on a wall console) and ex-
ternally connected to a power outlet/power supply.
It is not necessary to perform the leakage current measurement for
monitors mounted on the monitor trolley of the system if they are
part of the secondary circuit of the system power supply (e.g.,
SIREMOBIL Compact/Iso-C/POWERMOBIL/ARCADIS Varic/ARCA-
DIS Orbic).
Observe the leakage current measurement information in these in-
structions.
Electrical voltage!
Non-compliance can lead to severe injury and even death.
The leakage current measurement may be performed on
systems of protection class I only after the protective
conductor test has been passed.
Electrical voltage!
Non-compliance can lead to severe injury and even death.
No housing parts of the system may be touched during
direct measurement of the leakage current (measure-
ment setup according to Fig. 26). Third-person access to
the system must be prevented.
SPR2-230.101.01.02.02
08.05
Monitor 44cm, 100/120HZ
Page 21 of 26
CS PS 24
21