NOTE
Monitor 44cm, 100/120HZ
Evaluate the results by comparing the first measured value and the
values documented during preceding maintenance procedures or
safety checks to the measured value.
A sudden or unexpected increase in the measured values may in-
dicate that an error occurred in the primary circuit of the power
supply (damaged insulation, damage caused by water ingress or
humidity, defective interference suppressor, etc.) - even if the limit
value of 0.5 mA is not exceeded.
SPR2-230.101.01.02.02
08.05
Safety inspection
Page 22 of 26
CS PS 24
Siemens AG
Medical Solutions