General information
Pos. 1
Pos. 2
Pos. 3
If the direct measurement of the system leakage current is used
tem must be insulated during the measurement and must not be touched.
Fig. 4:
Pos. 1
Pos. 2
Pos. 3
WARNING
The system must be switched on during measurement. Measuring devices with auto-
mated measuring sequences must therefore be set to manual measurement.
The highest value must be entered in the system leakage current report.
This value must not exceed the permissible leakage current values according to DIN VDE
0751-1:2001-10, Table F.1, line "general leakage current information", of 0.5 mA.
Measure and record the current line voltage. If the measured line voltage deviates from
the nominal voltage, correct the measured value to the value corresponding to a mea-
surement at the nominal value of the line voltage. This is also to be documented.
Document the measuring procedure (differential measurement or direct measurement)
and the measuring device used (designation and serial number).
In the case of repeat measurements, the measured value is also to be evaluated.
Siemens AG
Medical Solutions
I.
System
Application part (not available)
Measurement setup (integrated into measuring device)
Measuring circuit for direct measurement of the system leakage current in compliance
with DIN VDE 0751-1:2001-10, Fig. C5 for protection class I.
System
Application part (not available)
Measurement setup (integrated into measuring device)
Electrical voltage!
Non-compliance can lead to severe injury and even death.
No housing parts of the system may be touched during direct mea-
surement of the system leakage current
Third-person access to the system must be prevented.
SPR2-230.101.01.02.02
08.05
(Fig. 4 / p.
(Fig. 4 / p.
Monitor 44cm, 100/120HZ
Page 13 of 26
CS PS 24
13
13), the sys-
13).