Siemens SITRANS LG240 Operating Instructions Manual page 68

Guided wave radar
Hide thumbs Also See for SITRANS LG240:
Table of Contents

Advertisement

Process conditions
For the process conditions, please also note the specifications on the type label. The lowest value
always applies.
The measurement error through the process conditions in the specified pressure and temperature
range is < 1 %.
Process pressure
Ʋ Clamp, flange version ≤ 2"/DN 50
Ʋ Clamp, flange version > 2"/DN 50
Vessel pressure relating to the flange
nominal pressure stage
Process temperature (Clamp or flange temperature)
A
80°C / 176°F
60°C / 140°F
45°C / 113°F
35°C / 95°F
30°C / 86°F
0°C / 32°F
-40°C
-40°F
-40°C / -40°F
Fig. 31: Ambient temperature - process temperature, standard version
A
Ambient temperature
B
Process temperature (depending on the seal material)
1
Aluminium housing
2
Plastic housing
3
Stainless steel housing, precision casting
4
Stainless steel housing, electropolished
Mechanical stress
Vibration resistance
Ʋ Rod probe
Shock resistance
Ʋ Rod probe
68
-0.5 ... +16 bar/-50 ... +1600 kPa (-7.3 ... +232 psig),
depending on the process fitting
-0.2 ... +16 bar/-20 ... +1600 kPa (-2.9 ... +232 psig),
depending on the process fitting
see supplementary instructions manual "Flanges ac-
cording to DIN-EN-ASME-JIS"
50°C
80°C
100°C
122°F
176°F
212°F
1 g with 5 ... 200 Hz according EN 60068-2-6 (vibration
at resonance) with rod length 50 cm (19.69 in)
25 g, 6 ms according to EN 60068-2-27 (mechanical
shock) with rod length 50 cm (19.69 in)
SITRANS LG240 - Operating Instructions
1
3
2
4
B
150°C
302°F
PBD-51041037

Advertisement