Custom White Point dialog

The Custom White Point dialog is accessed by clicking the Edit.. button in the White Point section of the Edit
Calibration Target dialog.
A custom White Point can be created by either specifying it as a Black Body radiator temperature in Kelvin, as
CIE (xy chromaticity) Coordinates, or by taking a measurement.
Values between 2500K and 10000K may be entered in the Black Body K field. The corresponding CIE xy
coordinates are automatically calculated and displayed for reference.
CIE xy chromaticity coordinates can be entered by selecting the CIE Coordinates radio button and entering the
x and y values. The corresponding correlated black body radiator temperature and Δuv (Delta-uv) values are
automatically calculated and displayed for reference. The Delta-uv value indicates how far the xy coordinates
entered are from the black body radiator locus.
A custom White Point may be measured using the calibration sensor, and automatically entered in the CIE
xy chromaticity coordinate fields by clicking the Measure button. This is useful for "cloning" the white point of
another display monitor.