 | NEC Display Wall Calibrator - UsEr's GUiDE

Summary window

The Summary window is accessed from the View menu.
This window shows the current layout configuration and connections, as well as detailed information about the
calibration results.
Connection Summary:
Port - Shows the Communications Port to the display, either the RS232 COM Port, or LAN IP Address.
Monitor ID - Shows the Monitor ID used to identify the display monitor.
Model - Shows the model of the display monitor.
Serial Number - Shows the serial number of the display monitor.
Status - Shows the current status of the display monitor in the measurement and calibration process.
Calibration Summary:
Uncalibrated Intensity – Shows the measured native Intensity (luminance) of the display when un-calibrated.
Uncalibrated White Point – Shows the native white point of the display when un-calibrated.
Uncalibrated Contrast Ratio – Shows the native contrast ratio of the display when un-calibrated.
Maximum Calibrated Intensity – Shows the calculated maximum Intensity that can be achieved by the display
using the current Target White Point setting. If the current Target Intensity method is set to Specific Level and
the calculated value is more than 5% lower than the Target, then a yellow warning icon is shown to warn that
the displays may not match in Intensity after calibration using the current Specific Level value, and a lower value
should be chosen.
Date Measured – Shows the date and time that the un-calibrated measurements were taken.
Operating Hours when Measured – Shows the number of hours the display has been in operation at the time
the un-calibrated measurements were taken.
Actual Calibrated Intensity – Shows the Intensity (luminance) of the display after it has been calibrated. If the
current Target is set to Maximum Possible Matching or Specific Level and the value is not within 10% of the
Target Intensity value, then a yellow alert icon is shown to warn that the display could not achieve the desired
Target Intensity value when calibrated.
Actual Calibrated White Point – Shows the white point that the display was adjusted to after calibration.
Date Calibrated - Shows the date and time when the calibration was performed.
Operating Hours when Calibrated – Shows the number of hours the display has been in operation when the
calibration was performed.