 | NEC Display Wall Calibrator - UsEr's GUiDE

Connection panel

assigned that must match that set on the display's OSD setting.
The communications settings and link to the display can be tested by clicking the Test button. This will attempt
to communicate with the display on the selected communications interface and Monitor ID.
Clicking the Update button will update the communications connection settings for the currently selected display
in the layout.
Note: If changes have been made, but the Update button has not been clicked, then the changes will be lost if
another display is selected in the layout.

Calibration Target panel

Contrast Ratio shows the Target contrast ratio.
This panel is used to modify the communications connection
settings for the currently selected display in the layout.
Connection types of RS232 or LAN (Ethernet) can be
selected.
Is using RS232, the RS232 COM port on the host PC used
to communicate with the displays can be selected. Currently
available COM ports on the host PC are listed in COM Port:.
Otherwise, if using LAN, the IP Address of the display can be
entered. See Chapter 8 for instructions on how to assign the
IP address for each display.
For both communications types, the Monitor ID of the display
must be selected and this must match the ID selected on the
display's OSD setting. If displays are daisy-chained together
using RS232, each display must have a unique Monitor ID
This panel shows the Calibration Target settings that will be
used when the displays are calibrated.
The Calibration Target settings can be modified by clicking
Edit... which will open the Edit Calibration Target dialog.
White Point shows the Target White Point.
Gamma shows the Target gamma or Tone Response Curve
type.
Intensity shows the Target display intensity adjustment
method and value.
Note: When using the Maximum Possible Matching method
for Intensity, the actual value used to calibrate all of the
displays will not be shown until all of the displays have been
measured, and the Target calculated.