Chapter 2 Operating Your Universal Counter

Using the Save and Recall Menus

Using the Save and Recall Menus

Overview of Save and Recall Functions

Utility
Local
Recall
Save &
Print
The combined use of the Save and Recall functions allows you to save and
recall the measurement setups which determine how the Counter
measures, analyzes, and reports answers. A setup can include the
measurement to be made, the precision desired, offset and scaling factor to
apply, limits to be met, and statistics to compute. Setups can also control
printing of hardcopy. Setups saved to memory registers are preserved
when the Counter is powered down or disconnected from a power source.
The Save and Recall functions enable faster and easier operation, fewer
operator errors, and minimum training.
The Save and Recall keys allow you to:
1
For firmware revisions 3622 and above, RECALL 0 is stored in volatile memory. RECALL 0 is not preserved
if you cycle power.
2-52
create and save up to 20 different instrument measurement setups
into internal non-volatile memory (SAVE:).
recall up to 20 user-selected measurement setups that were previously
saved (RECALL 1, 2, 3, ... 20). With a total of 21 save/recall
registers, there are 20 registers available to you for save/recall
operations. The Counter automatically saves the current setup to
register 0 (RECALL 0) before executing a recall function.
delete any of the saved setups (UNSAVE:).